3. Rubber and Plastics Cycle
The Rubber and Plastics Cycle is used to sterilize up to 0.4 kg (0.9 lbs) of solid or hollow
unwrapped instruments constructed of metal
or the materials listed in Section 4.4 Preparing
and Loading Instruments.
To select the Rubber and Plastics Cycle, press
the Rubber and Plastics Cycle button, then
press the START button. The sterilization
temperature in the cassette is 121˚C (250 ˚C)
and for the holding time is 35 minutes for USA
and 15 minutes for all other countries.
4. Heavy Duty Unwrapped Cycle
The Heavy Duty Cycle is used to sterilize larger
loads of unwrapped metal instruments weighing up
to 1.5 kg (3.3 lbs). Dental hand pieces can be
sterilized in this cycle. Using the Statim 5000
Extended Cassette (order no. 01-104104), rigid
endoscopes may be sterilized in this cycle.
To select the Heavy Duty Unwrapped Cycle, press
the Heavy Duty Unwrapped Cycle button, then press the START button.
The sterilization temperature in the cassette is 132˚ C (270˚ F) for USA and
134˚ C (273˚ F) for all other countries, and the holding time is 6 minutes.
4. Instructions For Use
A
B
C
D
E
F
A Warming up
B Conditioning
C Pressurizing
D Sterilizing
E Venting
F Air Drying
A
B
C
D
E
F
A
Warming up
B Conditioning
C Presurizing
D Sterilizing
E Venting
F Air Drying