4. Instructions For Use
1. Unwrapped Cycle
The Unwrapped Cycle is used to sterilize light loads
(less than 0.5 kg or 1.1 lbs) of solid metal instruments
such as pliers, burrs, scalers, and forceps.
To select the Unwrapped Cycle, press the
Unwrapped Cycle button, then press the START
button. The sterilization temperature in the cassette
is 132˚ C (270˚ F) for USA and 134˚ C (273˚ F) for all
other countries, and the holding time is 3.5 minutes. See Section 4.2 Statim 5000
Cassette and Section 4.4 Preparing and Loading Instruments before running this cycle.
2. Wrapped Cycle
The Wrapped Cycle is used to sterilize up to
1.5 kg (3.3 lbs) of solid and hollow metal
instruments which have been sealed in paper /
paper, or paper / plastic autoclave bags. Dental
hand pieces may be sterilized in this cycle.
To select the Wrapped Cycle, press the Wrapped Cycle button, then press the START
button. The sterilization temperature in the cassette is 132˚ C (270˚ F) for USA and 134˚
C (273˚ F) for all other countries, and the holding time is 6 minutes. See section 4.2
Statim 5000 Cassette and Section 4.4 Preparing and Loading Instruments before
running this cycle.
If wrapped instruments are intended for storage, the wraps must be dry when the
cassette is removed from the unit and opened.
A
B
C
D
E
F
A
Warming up
B C
onditioning
C Pressurizing
D
Sterilizing
E V
enting
F
Air Drying
3.5 MIN.*
A
B
C
D
E
F
A
Warming up
B Conditioning
C Pressurizing
D Sterilizing
E Venting
F Air Drying
23