4. Instructions For Use
2. Wrapped Cycle
The Wrapped Cycle is used to sterilize up
to 1.0 kg (2.2 lbs) of solid and hollow
metal instruments which have been
sealed in paper / paper, or paper / plastic
autoclave bags. Dental handpieces may
be sterilized in this cycle.
To select the Wrapped Cycle, press the
Wrapped Cycle button, then press the
START button.
The sterilization temperature in the cassette is 134˚C (273˚F) and the holding time is 10
minutes. See Section 4.1 Statim 2000 Cassette, and Section 4.4 Preparing and
Loading Instruments before running this cycle.
If wrapped instruments are intended for storage, the wraps must be dry when the
cassette is removed from the unit and opened.
3. Rubber and Plastics Cycle
The Rubber and Plastics Cycle is used to
sterilize up to 0.4 kg (0.9 lbs) of solid
unwrapped instruments constructed of
metal or the materials listed in Section
4.4 Preparing and Loading Instruments.
To select the Rubber and Plastics Cycle,
press the Rubber and Plastics Cycle
button, then press the START button.
The sterilization temperature in the cassette is 121˚C / 250˚F and the holding time is
15 minutes.
A
B
C
D
E
F
A
Warming up
B Conditioning
C Pressurizing
D Sterilizing
E Venting
F Air Drying