19
P1
Thermocouple
Microswitch
Solenoid
Valve
Air
Compressor
Steam
Generator/Boiler
Invensys
Water
Quality
Sensor
Thermocouple
Heater
Double Thermal Fuse
Reservoir
FWR
J1-1
J1-4
J1-5
J1-6
J1-7
J1-8
J1-9
J1-10
J3-6
W1
J3-1
J4-6
J4-1
P1
P3
P4
F1
F4
F2
F3
P2
Pump
Blue
Red
Black
White
Shield
Yellow
Red
Red
Yellow
Shield
Power
Switch
AC Receptical
White
Black
White
White
White
Black
White
Black
Black
White
Line
Neut.
Line
Neut.
Line
Neut.
Line
Neut.
GND
Chassis
Gnd
+
-
+
-
CHAMBER
BOILER
J5 J6
20
18
16
Power
Cycle
Over-Temp.
Switch
Blue
Blue
Float
(
Not Shown
)
#14 AWG
#14 AWG
#14
AWG
#14
AWG
Unmarked wires are #22 AWG
Wires used in this
terminal block need
to be solder tinned,
or an insulation
ferrule be used.
CR5
CR7
CR10
CR6
Green/Yellow
Green/Yellow
9.0 – REFERENCE BLOCKS
(continued)
FIGURE 9.8 - ELECTRICAL DIAGRAM
Keypad Chart Connect to P4
Pin # 1 2 3 4 5 6 7
Function Stop Start Hemostat Hand Closed Open Common
Scissors Scissors
Device Activation Chart (Jumpering)
Pump P1-16 to J3-5 (GND)
Solenoid P1-18 to J3-5 (GND)
Air Compressor P1-20 to J3-5 (GND)
Diagnostic LED Reference
Data Chart
Steam Generator/Boiler Heater: -11 Ω (J1-3 & J1-4)
Solenoid Coil (FWR): 3.25 - 3.96 Meg Ω
Thermocouples: ~10 Ω @ Room Temp. J5 Cassette; J6 Boiler
Thermal Fuses: Open @ 216˚C
Fuse Chart
Earlier Style Later Style
Replacement Fuse 2000 & 5000 2000 & 5000
RPI Part #RPF060 (1/2A)
1/4" x 1-1/4" Glass F3 & F4
2000 Only
RPI Part #RPF061 (1A) In-Line Pump
1/4" x 1-1/4" Glass (Not Shown)
2000 Only
RPI Part #RPF071 (15A) F1 & F2 F2 & F3
1/4" x 1-1/4" Ceramic Line Voltage Line Voltage
RPI Part #RPF367 (2A)
5mm x 20mm Glass F1
RPI Part #RPF368
5mm x 20mm Glass F2
CR5 Steam Generator/Boiler
CR6 Air Compressor
CR7 Solenoid Valve
CR10 Pump
Diagnostic LED's will be illumi-
nated when their corresponding
device is switched ON
If the Air Compressor, Solenoid Valve, Pump or Steam Generator/Boiler
activates, but does not turn off at the appropriate time, the Main PCB Board
is bad and must be replaced.